| Product Name: UTX750B X ray fluorescence spectrometer Order number: Optional accessories: no Inquiry: Detailed description One, application field UTX750B is a versatile X ray fluorescence spectrometer, suitable for: 1, RoHS detection The EU RoHS directive is restricted by the Cr, Hg, Pb, Cd, Br, Cl, I and other elements of the detection 2, coating testing Suitable for testing the larger size of the hardware, connectors, electrical connector plating, plating, plastics, PCB, ABS magnesium alloy electroplating. Particularly suitable for thin Au, Ag, Pd and other precious metal plating. Hardware, connectors, electrical connector plating Au, Ag, Sn, Ni, Cu, Zn, Cr, NiP, SnPb, SnCu, ZnNi, AuPdNi such as alloy; PCB Au, Ag Cu, electroplating, Ni, NiP, SnPb alloy; Plastic, ABS, magnesium alloy electroplating Au, Ag, Sn, Ni, Cu, Zn, Cr, NiP etc.. 3, plating bath analysis Electroplating solution of Au, Ag, Cu, Ni, Cr, Zn, Sn concentration was detected, and the impurity elements Ag, Cu, Ni, Cr, Zn, Sn, Fe, Co, Mn, Sb analysis. 4, alloy analysis Stainless steel in Fe, Cr, Ni, Mo and other components Alloy steel in Fe, Ti, V, Cr, Mn, Co, Ni, Cu, Zn, Nb, Mo, Sn, Sb, W, Pb and other components Copper alloy in Cu, Zn, Fe, Cr, Ni, Cu, Zn, Sn, Sb, W, Pb, and other ingredients Zinc alloy in Zn, Cu, Fe, Cr, Mn, Ni, Cu, Zn, Sn, Sb, Pb, W, and other ingredients Solder in Sn, Fe, V, Cr, Ni, Cu, Zn, Nb, Sb, W, Pb and other components 5, precious metal identification Gold jewelry testing purity and impurity Ag, Ni, Cu, Zn, Cd, Y, Zr, Rh, Pd analysis Platinum jewelry testing purity and impurity Ag, Ni, Cu, Zn, Cd, Y, Zr, Rh, Pd, Au analysis Palladium jewelry testing purity and impurity Ag, Ni, Cu, Zn, Cd, Y, Zr analysis Silver jewelry testing purity and impurity Fe, Ni, Cu, Zn, Cd, Sn analysis 6, food and safety inspection The daily food trace elements Fe, Ca, Zn, Cu, Pb, Cd, Ni analysis Analysis of the effective components in various supplements such as Ca oral liquid on the content of Ca, Zn oral liquid on Zn content Tea discrimination : according to the different origin of the trace elements in tea of different tea varieties, discrimination and authenticity 7, environment detection Gasoline, diesel, kerosene and other content in Pb, As, Cd and other elements of the detection Analysis of water quality analysis, water, sewage, river water, groundwater, drinking water Fe, Ti, V, Cr, Mn, Co, Ni, Cu, Zn, Nb, Mo, Sn, Sb, W, Pb, Hg, Cd and other elements Two, product characteristics 1, from Al to U rapid element analysis, metal plating thickness detection In 2, the detection sensitivity of 1ppm 99.99%; 0.005um to 35um coating In 3, 15-20 times CCD color system was used for sample image observation 4, in a variety of shapes and specifications of the collimator can be chosen In 5, a variety of filtering mechanism, a variety of specifications of filters for different samples In 6, high resolution of SIPIN detector, improve the measurement sensitivity In 7, the detector of electric refrigeration, without liquid nitrogen In 8, large sample chamber with precision regulating electric sample table, can measure large to small samples of various sizes In 9, FP basic algorithm software In 10, the flexible application of analysis software 11, simple installation, fast instrument Three, technical indicators 
| High-voltage power supply |   |  
| Input voltage | 24VDC±10% |   |  
| Output voltage | 0-50kV |   |  
| Output current | 0-2.0mA |   |  
| Output precision | 0.01% |   |  
| Remarks | Software automatic detection function |   |  
| X ray tube |   |  
| X ray tube target | W target, Rh, Mo, Ag, Cr target optional |   |  
| X ray tube window | Nylon window |   |  
| Electrical parameters | 50W、4-50kV,0-1000uA |   |  
| Cooling mode | Silicone grease cooling |   |  
| Service life | 15000-20000h |   |  
| Detector |   |  
| Detection range | Al19-U92 |   |  
| Resolution | FWHM 139eV |   |  
| The detection window | Be,Window area:5-25mm2 |   |  
| Thickness | 0.5-5mm |   |  
| Spectral processing system |   |  
| Processor type | Full digital 32-bit3 DSP |   |  
| The number of spectral channels | 256,512,1024,2048 choice |   |  
| Counting rate | More than 100000cps |   |  
| Energy range | 500eV-40960eV |   |  
| Dead time | <3% |   |  
|   | Light path system and sample observation |  
|   | Collimator system | φ8mmφ8mmφ8mmφ2mmφ0.5mm Switchable |  
|   | Minimum measuring point size | φ0.6mm |  
|   | Focusing system | Laser focus |  
|   | Filter | A variety of filter ( Ni, Cu, Mo, Al, Ti ) to choose |  
|   | Sample observation system | The 3000000 pixel high-definition CCD camera, 20 times optical zoom. |  
|   | RoHS measuring index | Analyses of 1ppm-99.99% Precision : the relative standard deviation is less than 5% ( content is greater than 500ppm ) Accuracy: the relative error less than 5% ( content is greater than 500ppm ) |  
|   | Thickness indicator | Analysis of range: 0.01-35um Precision and accuracy: the first layer of the relative standard deviation is less than 5% Second layer relative standard deviation less than 10% Third layer relative standard deviation less than 15% |  
|   | Plating bath analysis | Analysis of range: 0.01g-1000g / L Precision : the relative standard deviation is less than 5%. Accuracy: the relative error less than 5% |  
| Sample room |   |  
| Sample room interior space(mm) | (W x H x D):450×90×400mm |   |  
| Sample size | (宽×深):250×275mm |   |  
| Sample table bearing weight | ≤5kg |   |  
| The test sample size | W x H x D:440×80×390mm。 |   |  
| Sample table control | Automatic 3D X-Y-Z sample stage |   |  
| Work bench mobile | X=200±5、Y=200±5、Z=80±5mm |   |  
| Worktable moving precision | 0.005mm |   |  
|   |  Computer and software |   |  
|   | Control computer | Lenovo brand QitianM7150, 1GB memory, 160G hard disk |   |  
|   | Monitor | Lenovo 17 inch LCD display |   |  
|   | Operating system | WindowsXP、OEM simplified Chinese Edition |   |  
|   | Report results | Word、Excel format |   |  
|   | Software | RoHS detection software, coating thickness software, element analysis software  |   |  
|   | Types of software | Fundamental parameter method of FP software |   |  
|   | Software language | Simplified Chinese, traditional Chinese, English, other |   |  
|   | Spectrum display | Automatic peak qualitative, KLM markers, spectral overlapping comparison |   |  
|   | Spectral processing | Total intensity, net intensity, background intensity; the whole area, net area, much element fitting, Gauss fitting; logarithmic spectrum |   |  
|   | Data analysis | The basic parameters, basic pattern, without standard sample test, peak position correction, substrate amendment |   |  
| Weight and size |   |  
| The main body of the device size(mm) | W 580×H 635×D 770mm |   |  
| Body weight | About 80kg |   |  
| The environment of the use |   |  
| Working environment | Temperature 15-250C,Humidity  40-70%RH |   |  
| Power supply system | Single-phase  220V±10%,Working voltage in allowed range |   |  
| Other | 1.      Don’t close to generate strong magnetic field, electric field, high frequency device 2.     Reducing vibration 3.      Less dust, low humidity, no corrosive gas 4.      No direct sunlight 5.      As the earthquake countermeasures, to consider the fixed device |   |  
|  |  |  |  |  |  |  |  |