Product Name: X ray fluorescence spectrometer UTX750
Order number:
Optional accessories: no
Inquiry:
Detailed description
One, Product brief introduction
UTX750 is a versatile X ray fluorescence spectrometer, suitable for:
In 1, RoHS detection
The EU RoHS directive is restricted by the Cr, Hg, Pb, Cd, Br, Cl, I and other elements of the detection
2, plating bath analysis
Electroplating solution of Au, Ag, Cu, Ni, Cr, Zn, Sn concentration was detected, and the impurity elements Ag, Cu, Ni, Cr, Zn, Sn, Fe, Co, Mn, Sb analysis.
3, alloy analysis
Stainless steel in Fe, Cr, Ni, Mo and other components
Alloy steel in Fe, Ti, V, Cr, Mn, Co, Ni, Cu, Zn, Nb, Mo, Sn, Sb, W, Pb and other components
Copper alloy in Cu, Zn, Fe, Cr, Ni, Cu, Zn, Sn, Sb, W, Pb, and other ingredients
Zinc alloy in Zn, Cu, Fe, Cr, Mn, Ni, Cu, Zn, Sn, Sb, Pb, W, and other ingredients
Solder in Sn, Fe, V, Cr, Ni, Cu, Zn, Nb, Sb, W, Pb and other components
4, precious metal identification
Gold jewelry testing purity and impurity Ag, Ni, Cu, Zn, Cd, Y, Zr, Rh, Pd analysis
Platinum jewelry testing purity and impurity Ag, Ni, Cu, Zn, Cd, Y, Zr, Rh, Pd, Au analysis
Palladium jewelry testing purity and impurity Ag, Ni, Cu, Zn, Cd, Y, Zr analysis
Silver jewelry testing purity and impurity Fe, Ni, Cu, Zn, Cd, Sn analysis
5, food and safety inspection
The daily food trace elements Fe, Ca, Zn, Cu, Pb, Cd, Ni analysis
Analysis of the effective components in various supplements such as Ca oral liquid on the content of Ca, Zn oral liquid on Zn content
Tea discrimination : according to the different origin of the trace elements in tea of different tea varieties, discrimination and authenticity
6, environment detection
Gasoline, diesel, kerosene and other content in Pb, As, Cd and other elements of the detection
Analysis of water quality analysis, water, sewage, river water, groundwater, drinking water Fe, Ti, V, Cr, Mn, Co, Ni, Cu, Zn, Nb, Mo, Sn, Sb, W, Pb, Hg, Cd and other elements
Two, product characteristics
1, the detection sensitivity of 2ppm to 99.99%
2, 15-20 times CCD color system was used for sample image observation
3, in a variety of shapes and specifications of the collimator can be chosen
4, a variety of filtering mechanism, a variety of specifications of filters for different samples
5, high resolution of SIPIN detector, improve the measurement sensitivity
6, the detector of electric refrigeration, without liquid nitrogen
7, large sample storehouse, can measure large to small samples of various sizes
8, FP basic algorithm software
9, the flexible application of analysis software
10, simple installation, fast instrument
Three, technical indicators
High-voltage power supply |
|
Input voltage |
24VDC±10% |
|
Output voltage |
0-50kV |
|
Output current |
0-2.0mA |
|
Output precision |
0.01% |
|
Remarks |
Software automatic detection function |
|
X ray tube |
|
X ray tube target |
W target, Rh, Mo, Ag, Cr target optional |
|
X ray tube window |
Nylon window |
|
Electrical parameters |
50W、4-50kV,0-1000uA |
|
Cooling mode |
Electric refrigeration |
|
Service life |
15000-20000h |
|
Detector |
|
Detection range |
Al19-U92 |
|
Resolution |
FWHM 139eV |
|
The detection window |
Be,Window area:5-25mm2 |
|
Thickness |
0.5-5mm |
|
RoHS measuring index |
|
Analysis range |
1ppm—99.99% |
|
Precision |
The relative standard deviation below 5% ( content is greater than 500ppm ) |
|
Degree of accuracy |
Relative error below 5% ( content is greater than 500ppm ) |
|
Spectral processing system |
|
Processor type |
Full digital 32-bit3 DSP |
|
The number of spectral channels |
256,512,1024,2048 choice |
|
Counting rate |
More than 100000cps |
|
Energy range |
500eV-40960eV |
|
Dead time |
<3% |
|
|
Light path system and sample observation |
|
Collimator system |
φ8mm*3φ2mmφ0.5mm Switchable |
|
Focusing system |
Laser focus |
|
Filter |
A variety of filter ( Ni, Cu, Mo, Al, Ti ) to choose |
|
Sample observation system |
The 3000000 pixel high-definition CCD camera, 20 times optical zoom. |
Sample room |
|
Sample room interior space(mm) |
(W x H x D):450×160×400mm |
|
Sample size |
(宽×深):250×275mm |
|
Sample table bearing weight |
≤5kg |
|
The test sample size |
W x H x D:440×150×390mm。 |
|
Sample table control |
Manual sample stage |
|
Work bench mobile |
X=200±5、Y=200±5、Z=160±5mm |
|
|
Computer and software |
|
|
Control computer |
Lenovo brand T4900V, 512MB memory, 80G hard disk |
|
|
Monitor |
Lenovo 17 inch LCD display |
|
|
Operating system |
WindowsXP、OEM simplified Chinese Edition |
|
|
Report results |
Word、Excel format |
|
|
Software |
RoHS detection software, coating thickness software, element analysis software |
|
|
Types of software |
Fundamental parameter method of FP software |
|
|
Software language |
Simplified Chinese, traditional Chinese, English, other |
|
|
Spectrum display |
Automatic peak qualitative, KLM markers, spectral overlapping comparison |
|
|
Spectral processing |
Total intensity, net intensity, background intensity; the whole area, net area, much element fitting, Gauss fitting; logarithmic spectrum |
|
|
Data analysis |
The basic parameters, basic pattern, without standard sample test, peak position correction, substrate amendment |
|
Weight and size |
|
The main body of the device size(mm) |
W 580×H 635×D 770mm |
|
Body weight |
About 60kg |
|
The environment of the use |
|
Working environment |
Temperature 15-250C,Humidity 40-70%RH |
|
Power supply system |
Single-phase 220V±10%,Working voltage in allowed range |
|
Other |
1. Don’t close to generate strong magnetic field, electric field, high frequency device
2. Reducing vibration
3. Less dust, low humidity, no corrosive gas
4. No direct sunlight
5. As the earthquake countermeasures, to consider the fixed device |
|
|
|
|
|
|
|
|